Compatibility of medications during ...
Type de document :
Article dans une revue scientifique
PMID :
URL permanente :
Titre :
Compatibility of medications during multi-infusion therapy: A controlled in vitro study on a multilumen infusion device
Auteur(s) :
Perez, Maxime [Auteur]
Decaudin, Bertrand [Auteur]
Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Foinard, Aurelie [Auteur]
Barthelemy, Christine [Auteur]
Debaene, Bertrand [Auteur]
Lebuffe, Gilles [Auteur]
Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Odou, Pascal [Auteur]
Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Decaudin, Bertrand [Auteur]
Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Foinard, Aurelie [Auteur]
Barthelemy, Christine [Auteur]
Debaene, Bertrand [Auteur]
Lebuffe, Gilles [Auteur]
Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Odou, Pascal [Auteur]
Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Titre de la revue :
Anaesthesia, Critical Care & Pain Medicine
Nom court de la revue :
Anaesth. Crit. Care Pain Med.
Numéro :
34
Pagination :
83-88
Date de publication :
2015
Discipline(s) HAL :
Sciences du Vivant [q-bio]
Résumé en anglais : [en]
OBJECTIVE: Drug incompatibilities can jeopardize the safety and effectiveness of intravenous drug therapies, especially in the field of anaesthesia and intensive care. Patients receive many drugs simultaneously through ...
Lire la suite >OBJECTIVE: Drug incompatibilities can jeopardize the safety and effectiveness of intravenous drug therapies, especially in the field of anaesthesia and intensive care. Patients receive many drugs simultaneously through limited venous accesses. This study was designed to confirm the impact of a multilumen infusion device on the occurrence of known physical drug incompatibilities. STUDY DESIGN: In vitro laboratory work. METHODS: Two infusion devices were studied: a standard single-lumen set and a multilumen infusion access device (Edelvaiss Multiline-8, Doran International). Up to six drugs were infused simultaneously: three acidic solutions of midazolam, amiodarone and dobutamine, and three alkaline solutions of furosemide, pantoprazole and amoxicillin/clavulanate. Saline, Ringer' solution and 5% dextrose were used as hydration vehicles with an infusion rate initially set at 100 mL/h and with stepwise decreases of 10 mL/h until precipitation. Two methods were used to highlight physical drug compatibility according to the European Pharmacopoeia: visual inspection of the extension set and an obscured-light sub-visible particle count test of infusions. The lowest infusion rate value for vehicle infusion to satisfy the two tests in all trials is reported for each infusion device. RESULTS: The standard set did not satisfy the test in 82% of the assessed drug combinations. The Edelvaiss Multiline-8 was able to prevent the occurrence of drug incompatibilities in 49% of the drug combinations tested. This device is therefore advantageous, especially when simultaneously infusing two or four incompatible drugs. CONCLUSIONS: Infusion device characteristics have an impact on physical drug incompatibilities. Our results confirm that the Edelvaiss Multiline-8 device prevents physical drug incompatibilities under specified conditions.Lire moins >
Lire la suite >OBJECTIVE: Drug incompatibilities can jeopardize the safety and effectiveness of intravenous drug therapies, especially in the field of anaesthesia and intensive care. Patients receive many drugs simultaneously through limited venous accesses. This study was designed to confirm the impact of a multilumen infusion device on the occurrence of known physical drug incompatibilities. STUDY DESIGN: In vitro laboratory work. METHODS: Two infusion devices were studied: a standard single-lumen set and a multilumen infusion access device (Edelvaiss Multiline-8, Doran International). Up to six drugs were infused simultaneously: three acidic solutions of midazolam, amiodarone and dobutamine, and three alkaline solutions of furosemide, pantoprazole and amoxicillin/clavulanate. Saline, Ringer' solution and 5% dextrose were used as hydration vehicles with an infusion rate initially set at 100 mL/h and with stepwise decreases of 10 mL/h until precipitation. Two methods were used to highlight physical drug compatibility according to the European Pharmacopoeia: visual inspection of the extension set and an obscured-light sub-visible particle count test of infusions. The lowest infusion rate value for vehicle infusion to satisfy the two tests in all trials is reported for each infusion device. RESULTS: The standard set did not satisfy the test in 82% of the assessed drug combinations. The Edelvaiss Multiline-8 was able to prevent the occurrence of drug incompatibilities in 49% of the drug combinations tested. This device is therefore advantageous, especially when simultaneously infusing two or four incompatible drugs. CONCLUSIONS: Infusion device characteristics have an impact on physical drug incompatibilities. Our results confirm that the Edelvaiss Multiline-8 device prevents physical drug incompatibilities under specified conditions.Lire moins >
Langue :
Anglais
Audience :
Internationale
Vulgarisation :
Non
Établissement(s) :
Université de Lille
CHU Lille
CHU Lille
Collections :
Équipe(s) de recherche :
Innovation/évaluation des médicaments injectables
Modélisation biopharmaceutique et pharmacocinétique
Innovation/évaluation des dispositifs médicaux de perfusion
Modélisation biopharmaceutique et pharmacocinétique
Innovation/évaluation des dispositifs médicaux de perfusion
Date de dépôt :
2019-02-26T17:07:38Z