Femoral head diameter considerations for primary total hip arthroplasty
Auteur(s) :
Girard, Julien [Auteur]
Unité de Recherche Pluridisciplinaire Sport, Santé, Société (URePSSS) - ULR 7369
Unité de Recherche Pluridisciplinaire Sport, Santé, Société (URePSSS) - ULR 7369 - ULR 4488 [URePSSS]

Unité de Recherche Pluridisciplinaire Sport, Santé, Société (URePSSS) - ULR 7369
Unité de Recherche Pluridisciplinaire Sport, Santé, Société (URePSSS) - ULR 7369 - ULR 4488 [URePSSS]
Titre de la revue :
Orthopaedics & traumatology, surgery & research
Nom court de la revue :
Orthop. Traumatol.-Surg. Res.
Numéro :
101
Pagination :
S25-S29
Date de publication :
2015-02-01
ISSN :
1877-0568
Mot(s)-clé(s) en anglais :
Hip resurfacing
Dislocation
Large-diameter femoral head
Aseptic loosening
Dislocation
Large-diameter femoral head
Aseptic loosening
Discipline(s) HAL :
Sciences du Vivant [q-bio]
Résumé en anglais : [en]
The configuration of total hip arthroplasty (THA) implants has constantly evolved since they were first introduced. One of the key components of THA design is the diameter of the prosthetic femoral head. It has been well ...
Lire la suite >The configuration of total hip arthroplasty (THA) implants has constantly evolved since they were first introduced. One of the key components of THA design is the diameter of the prosthetic femoral head. It has been well established that the risk of dislocation is lower as the head diameter increases. But head diameter impacts other variables beyond joint stability: wear, cam-type impingement, range of motion, restoration of biomechanics, proprioception and groin pain. The introduction of highly cross-linked polyethylene and hard-on-hard bearings has allowed surgeons to implant large-diameter heads that almost completely eliminate the risk of dislocation. But as a result, cup liners have become thinner. With femoral head diameters up to 36 mm, the improvement in joint range of motion, delay in cam-type impingement and reduction in dislocation risk have been clearly demonstrated. Conversely, large-diameter heads do not provide any additional improvements. If an "ecologically sound" approach to hip replacement is embraced (e.g. keeping the native femoral head diameter), hip resurfacing with a metal-on-metal bearing must be carried out. The reliability of large-diameter femoral heads in the longer term is questionable. Large-diameter ceramic-on-ceramic bearings may be plagued by the same problems as metal-on-metal bearings: groin pain, squeaking, increased stiffness, irregular lubrication, acetabular loosening and notable friction at the Morse taper. These possibilities require us to be extra careful when using femoral heads with a diameter greater than 36 mm.Lire moins >
Lire la suite >The configuration of total hip arthroplasty (THA) implants has constantly evolved since they were first introduced. One of the key components of THA design is the diameter of the prosthetic femoral head. It has been well established that the risk of dislocation is lower as the head diameter increases. But head diameter impacts other variables beyond joint stability: wear, cam-type impingement, range of motion, restoration of biomechanics, proprioception and groin pain. The introduction of highly cross-linked polyethylene and hard-on-hard bearings has allowed surgeons to implant large-diameter heads that almost completely eliminate the risk of dislocation. But as a result, cup liners have become thinner. With femoral head diameters up to 36 mm, the improvement in joint range of motion, delay in cam-type impingement and reduction in dislocation risk have been clearly demonstrated. Conversely, large-diameter heads do not provide any additional improvements. If an "ecologically sound" approach to hip replacement is embraced (e.g. keeping the native femoral head diameter), hip resurfacing with a metal-on-metal bearing must be carried out. The reliability of large-diameter femoral heads in the longer term is questionable. Large-diameter ceramic-on-ceramic bearings may be plagued by the same problems as metal-on-metal bearings: groin pain, squeaking, increased stiffness, irregular lubrication, acetabular loosening and notable friction at the Morse taper. These possibilities require us to be extra careful when using femoral heads with a diameter greater than 36 mm.Lire moins >
Langue :
Anglais
Audience :
Internationale
Vulgarisation :
Non
Établissement(s) :
Univ. Littoral Côte d’Opale
Univ. Artois
Université de Lille
Univ. Artois
Université de Lille
Équipe(s) de recherche :
Activité Physique, Muscle, Santé (APMS)
Date de dépôt :
2019-09-24T07:26:28Z