Long-term comparative study of large-diameter ...
Document type :
Article dans une revue scientifique: Article original
PMID :
Permalink :
Title :
Long-term comparative study of large-diameter metal-on-metal bearings: Resurfacing versus total arthroplasty with large-diameter Durom™ bearing.
Author(s) :
Ridon, Pierre-Emmanuel [Auteur]
Putman, Sophie [Auteur]
Evaluation des technologies de santé et des pratiques médicales - ULR 2694 [METRICS]
Migaud, Henri [Auteur]
Marrow Adiposity & Bone Lab (MABLab) - ULR 4490
Marrow Adiposity & Bone Lab - Adiposité Médullaire et Os - ULR 4490 [MABLab]
Berton, Charles [Auteur]
Pasquier, Gilles [Auteur]
Marrow Adiposity & Bone Lab (MABLab) - ULR 4490
Marrow Adiposity & Bone Lab - Adiposité Médullaire et Os - ULR 4490 [MABLab]
Girard, Julien [Auteur]
Unité de Recherche Pluridisciplinaire Sport, Santé, Société (URePSSS) - ULR 7369
Unité de Recherche Pluridisciplinaire Sport, Santé, Société (URePSSS) - ULR 7369 - ULR 4488 [URePSSS]
Putman, Sophie [Auteur]
Evaluation des technologies de santé et des pratiques médicales - ULR 2694 [METRICS]
Migaud, Henri [Auteur]
Marrow Adiposity & Bone Lab (MABLab) - ULR 4490
Marrow Adiposity & Bone Lab - Adiposité Médullaire et Os - ULR 4490 [MABLab]
Berton, Charles [Auteur]
Pasquier, Gilles [Auteur]
Marrow Adiposity & Bone Lab (MABLab) - ULR 4490
Marrow Adiposity & Bone Lab - Adiposité Médullaire et Os - ULR 4490 [MABLab]
Girard, Julien [Auteur]
Unité de Recherche Pluridisciplinaire Sport, Santé, Société (URePSSS) - ULR 7369
Unité de Recherche Pluridisciplinaire Sport, Santé, Société (URePSSS) - ULR 7369 - ULR 4488 [URePSSS]
Journal title :
Orthopaedics & Traumatology: Surgery & Research
Abbreviated title :
Orthop Traumatol Surg Res
Publication date :
2019-06-10
ISSN :
1877-0568
HAL domain(s) :
Sciences du Vivant [q-bio]
English abstract : [en]
BACKGROUND: Short-term results in total hip arthroplasty (THA) with large-diameter metal-on-metal (MoM) bearings were encouraging, but high failure rates have been reported in the long term, notably implicating corrosion ...
Show more >BACKGROUND: Short-term results in total hip arthroplasty (THA) with large-diameter metal-on-metal (MoM) bearings were encouraging, but high failure rates have been reported in the long term, notably implicating corrosion due to modularity. Several studies compared resurfacing (to which modularity does not apply) versus large-diameter MoM THA; but, to our knowledge, none compared the same bearing in the two situations with more than 10 years' follow-up. We therefore conducted a retrospective case-control study, using a single cup model (Durom™, Zimmer, Warsaw, USA) for both resurfacing (R) and large-diameter THA, to determine the role of modularity in failure of large-diameter MoM bearings. The study compared (1) metallic ion levels, and (2) survival, functional scores and complications rates between R and THA. OBJECTIVE: Large-diameter MoM bearing failure implicates not bearing wear but head-neck junction modularity in larger-diameter MoM THA. METHODS: Eighty-three THAs and 90 Rs were included between February 2004 and March 2006. All patients had clinical and radiologic follow-up with chromium (Cr) and cobalt (Co) ion blood assay. RESULTS: In the THA group, 24 of the 83 patients (28.9%) underwent revision for adverse reaction to metal debris (ARMD), versus none in the R group. Ten-year all-cause survival was significantly better in R (97.7%; 95% CI, 96.2-99.2) than THA (67.1%; 95% CI, 60.9-73.3). Median blood ion level was higher in THA (with a difference between Co and Cr: 5.75μg/L (range, 3.82-19.2) versus 1.75μg/L (range, 1.34-2.94) respectively) than in R (no difference: 0.89μg/L (range, 0.67-2.89) and 1.07μg/L (range, 0.67-1.65) respectively). In the THA group, there were positive correlations between Co and Cr elevation and implant revision (both p<0.0001). Co/Cr ratio was significantly higher in THA (2.57) than R (0.88) (p<0.0001), and higher again in the 24 cases of THA revision (4.67). There was no significant difference in mean PMA score (THA: 17.08±1.82 (range, 7-18); R: 17.50±0.74 (range, 15-18)), whereas mean Oxford score was better in R (14.32±2.5 (range, 12-24)) than THA (18.17±8.05 (range, 12-42)) (p=0.02). CONCLUSIONS: The present study confirmed the incontrovertible implication of modularity in failure of large-diameter MoM THA, by analyzing the same bearing in THA and in resurfacing. Trunnionosis was observed in the 24 cases of revision, with the THA adaptation ring inducing serious metallic ion release (with dissociated Co/Cr ratio), accounting for the high rate of revision. METHODS: III, case-control study.Show less >
Show more >BACKGROUND: Short-term results in total hip arthroplasty (THA) with large-diameter metal-on-metal (MoM) bearings were encouraging, but high failure rates have been reported in the long term, notably implicating corrosion due to modularity. Several studies compared resurfacing (to which modularity does not apply) versus large-diameter MoM THA; but, to our knowledge, none compared the same bearing in the two situations with more than 10 years' follow-up. We therefore conducted a retrospective case-control study, using a single cup model (Durom™, Zimmer, Warsaw, USA) for both resurfacing (R) and large-diameter THA, to determine the role of modularity in failure of large-diameter MoM bearings. The study compared (1) metallic ion levels, and (2) survival, functional scores and complications rates between R and THA. OBJECTIVE: Large-diameter MoM bearing failure implicates not bearing wear but head-neck junction modularity in larger-diameter MoM THA. METHODS: Eighty-three THAs and 90 Rs were included between February 2004 and March 2006. All patients had clinical and radiologic follow-up with chromium (Cr) and cobalt (Co) ion blood assay. RESULTS: In the THA group, 24 of the 83 patients (28.9%) underwent revision for adverse reaction to metal debris (ARMD), versus none in the R group. Ten-year all-cause survival was significantly better in R (97.7%; 95% CI, 96.2-99.2) than THA (67.1%; 95% CI, 60.9-73.3). Median blood ion level was higher in THA (with a difference between Co and Cr: 5.75μg/L (range, 3.82-19.2) versus 1.75μg/L (range, 1.34-2.94) respectively) than in R (no difference: 0.89μg/L (range, 0.67-2.89) and 1.07μg/L (range, 0.67-1.65) respectively). In the THA group, there were positive correlations between Co and Cr elevation and implant revision (both p<0.0001). Co/Cr ratio was significantly higher in THA (2.57) than R (0.88) (p<0.0001), and higher again in the 24 cases of THA revision (4.67). There was no significant difference in mean PMA score (THA: 17.08±1.82 (range, 7-18); R: 17.50±0.74 (range, 15-18)), whereas mean Oxford score was better in R (14.32±2.5 (range, 12-24)) than THA (18.17±8.05 (range, 12-42)) (p=0.02). CONCLUSIONS: The present study confirmed the incontrovertible implication of modularity in failure of large-diameter MoM THA, by analyzing the same bearing in THA and in resurfacing. Trunnionosis was observed in the 24 cases of revision, with the THA adaptation ring inducing serious metallic ion release (with dissociated Co/Cr ratio), accounting for the high rate of revision. METHODS: III, case-control study.Show less >
Language :
Anglais
Peer reviewed article :
Oui
Audience :
Internationale
Popular science :
Non
Administrative institution(s) :
CHU Lille
Univ. Littoral Côte d’Opale
Univ. Artois
Université de Lille
Univ. Littoral Côte d’Opale
Univ. Artois
Université de Lille
Collections :
Research team(s) :
Activité Physique, Muscle, Santé (APMS)
Submission date :
2019-09-24T07:28:10Z
2021-05-20T13:09:46Z
2024-02-26T10:19:41Z
2021-05-20T13:09:46Z
2024-02-26T10:19:41Z
Files
- 1-s2.0-S1877056819301562-main.pdf
- Version éditeur
- Open access
- Access the document