Influence of wilson ce nter terminal on fetal electrocardiography acquisition
Author(s) :
Charlier, P. [Auteur]
Logier, Regis [Auteur]
Evaluation des technologies de santé et des pratiques médicales - ULR 2694 [METRICS]
Cabon, Sarah [Auteur]
Garabedian, Charles [Auteur]
Debarge, Veronique [Auteur]
Environnement périnatal et croissance - EA 4489 [EPS]
De Jonckheere, Julien [Auteur]
Environnement périnatal et croissance - EA 4489 [EPS]
Logier, Regis [Auteur]
Evaluation des technologies de santé et des pratiques médicales - ULR 2694 [METRICS]
Cabon, Sarah [Auteur]
Garabedian, Charles [Auteur]

Debarge, Veronique [Auteur]
Environnement périnatal et croissance - EA 4489 [EPS]
De Jonckheere, Julien [Auteur]

Environnement périnatal et croissance - EA 4489 [EPS]
Journal title :
Conference proceedings . ... Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual Conference
Abbreviated title :
Conf Proc IEEE Eng Med Biol Soc
Volume number :
2018
Pages :
5545-5548
Publication date :
2018-07-01
ISSN :
1557-170X
HAL domain(s) :
Sciences du Vivant [q-bio]
English abstract : [en]
The acquisition of a standard 10-leads electrocardiography (ECG) is performed using the Wilson Center Terminal (WCT) reference with a normalized electrode positioning. However, in the case of non-invasive fetal ECG (fECG) ...
Show more >The acquisition of a standard 10-leads electrocardiography (ECG) is performed using the Wilson Center Terminal (WCT) reference with a normalized electrode positioning. However, in the case of non-invasive fetal ECG (fECG) acquisition, there is no standardization on the positioning of the electrodes on the abdomen and many authors suggest an acquisition with or without a WCT. The use of the WCT for the acquisition of the fetal heart rate (FHR) is not justified. The objective of this paper is to quantify the influence of this reference compared to a direct measurement. For this purpose, we developed a device allowing the acquisition of the fECG and compared the two configurations on 6 volunteer pregnants. The noise levels and the fetal QRS morphology were compared, showing no superiority of the WCT acquisition compared to a direct differential measurement.Show less >
Show more >The acquisition of a standard 10-leads electrocardiography (ECG) is performed using the Wilson Center Terminal (WCT) reference with a normalized electrode positioning. However, in the case of non-invasive fetal ECG (fECG) acquisition, there is no standardization on the positioning of the electrodes on the abdomen and many authors suggest an acquisition with or without a WCT. The use of the WCT for the acquisition of the fetal heart rate (FHR) is not justified. The objective of this paper is to quantify the influence of this reference compared to a direct measurement. For this purpose, we developed a device allowing the acquisition of the fECG and compared the two configurations on 6 volunteer pregnants. The noise levels and the fetal QRS morphology were compared, showing no superiority of the WCT acquisition compared to a direct differential measurement.Show less >
Language :
Anglais
Audience :
Internationale
Popular science :
Non
Administrative institution(s) :
CHU Lille
Université de Lille
Université de Lille
Submission date :
2019-12-09T16:48:33Z
2021-05-14T09:15:42Z
2021-05-14T09:15:42Z