Evaluation of Specific Administering Devices for Antineoplastic Drugs Compounded in Syringes in Paediatry: Methodology Proposal.
Auteur(s) :
Lefebvre, Maximilien [Auteur]
Simon, Nicolas [Auteur]
Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Vasseur, Michele [Auteur]
Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Sidikou, Ousseini [Auteur]
Barthelemy, Christine [Auteur]
Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Decaudin, Bertrand [Auteur]
Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Odou, Pascal [Auteur]
Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Simon, Nicolas [Auteur]
Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Vasseur, Michele [Auteur]
Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Sidikou, Ousseini [Auteur]
Barthelemy, Christine [Auteur]
Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Decaudin, Bertrand [Auteur]
Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Odou, Pascal [Auteur]
Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Titre de la revue :
Journal of oncology pharmacy practice
Nom court de la revue :
J. Oncol. Pharm. Pract.
Pagination :
178155217733323
Date de publication :
2017
Discipline(s) HAL :
Sciences du Vivant [q-bio]
Résumé en anglais : [en]
Purpose The aims of this study were to propose a simple methodology to assess the rinsing volume of syringe extension sets and to compare several marketed devices. Methods A UV-spectrophotometry assay using quinine ...
Lire la suite >Purpose The aims of this study were to propose a simple methodology to assess the rinsing volume of syringe extension sets and to compare several marketed devices. Methods A UV-spectrophotometry assay using quinine hydrochloride as drug substitute was developed. Quinine concentration ranged from 20 to 200 µg/ml. The assay was validated with the accuracy profile method and tested on five different assemblies (device+extension sets) with different dead-space volumes (1.28-2.80 ml) and at two different quinine concentrations (0.3 and 8.0 mg/ml). Rinsing was performed stepwise with water for injection until reaching an undetectable quinine concentration. After fitting the data with a Weibull model, assemblies were compared with an ANOVA performed on ranks (GraphPad, La Jolla, USA). Results The within-day and between-day precision ranges were 0.39-0.81 and 0.48-0.84%, respectively. The lower limit of quantification was 4.26 µg/ml. The volume required to completely rinse the infusion line was different according to the initial drug concentration and to the device assessed: from 6 to 10 ml for a low quinine concentration and from 7 to 17 ml for a high quinine concentration. Conclusion This study shows that a simple, cheap and easy-to-use methodology may be used to assess the rinsing volume of syringe extension sets. The rinsing volume is different according to the tested device.Lire moins >
Lire la suite >Purpose The aims of this study were to propose a simple methodology to assess the rinsing volume of syringe extension sets and to compare several marketed devices. Methods A UV-spectrophotometry assay using quinine hydrochloride as drug substitute was developed. Quinine concentration ranged from 20 to 200 µg/ml. The assay was validated with the accuracy profile method and tested on five different assemblies (device+extension sets) with different dead-space volumes (1.28-2.80 ml) and at two different quinine concentrations (0.3 and 8.0 mg/ml). Rinsing was performed stepwise with water for injection until reaching an undetectable quinine concentration. After fitting the data with a Weibull model, assemblies were compared with an ANOVA performed on ranks (GraphPad, La Jolla, USA). Results The within-day and between-day precision ranges were 0.39-0.81 and 0.48-0.84%, respectively. The lower limit of quantification was 4.26 µg/ml. The volume required to completely rinse the infusion line was different according to the initial drug concentration and to the device assessed: from 6 to 10 ml for a low quinine concentration and from 7 to 17 ml for a high quinine concentration. Conclusion This study shows that a simple, cheap and easy-to-use methodology may be used to assess the rinsing volume of syringe extension sets. The rinsing volume is different according to the tested device.Lire moins >
Langue :
Anglais
Audience :
Internationale
Vulgarisation :
Non
Établissement(s) :
Université de Lille
CHU Lille
CHU Lille
Collections :
Équipe(s) de recherche :
Innovation/évaluation des médicaments injectables
Modélisation biopharmaceutique et pharmacocinétique
Innovation/évaluation des dispositifs médicaux de perfusion
Modélisation biopharmaceutique et pharmacocinétique
Innovation/évaluation des dispositifs médicaux de perfusion
Date de dépôt :
2019-02-26T17:07:05Z
2019-07-05T12:31:29Z
2019-07-05T12:31:29Z