Effect of insulin infusion line on glycaemic variability in a perioperative high dependency unit (HDU): a prospective randomised controlled trial
Auteur(s) :
GENAY, Stéphanie [Auteur]
Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Decaudin, Bertrand [Auteur]
Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Ethgen, Sabine [Auteur]
Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Alluin, Arnaud [Auteur]
Babol, Elodie [Auteur]
Labreuche, Julien [Auteur]
Evaluation des technologies de santé et des pratiques médicales - ULR 2694 [METRICS]
Behal, Helene [Auteur]
Evaluation des technologies de santé et des pratiques médicales - ULR 2694 [METRICS]
Vantyghem, Marie-Christine [Auteur]
Odou, Pascal [Auteur]
Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Lebuffe, Gilles [Auteur]
Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]

Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Decaudin, Bertrand [Auteur]

Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Ethgen, Sabine [Auteur]

Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Alluin, Arnaud [Auteur]
Babol, Elodie [Auteur]
Labreuche, Julien [Auteur]
Evaluation des technologies de santé et des pratiques médicales - ULR 2694 [METRICS]
Behal, Helene [Auteur]

Evaluation des technologies de santé et des pratiques médicales - ULR 2694 [METRICS]
Vantyghem, Marie-Christine [Auteur]

Odou, Pascal [Auteur]

Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Lebuffe, Gilles [Auteur]

Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Titre de la revue :
Annals of intensive care
Nom court de la revue :
Ann. Intensive Care
Numéro :
7
Date de publication :
2017
Discipline(s) HAL :
Sciences du Vivant [q-bio]
Résumé en anglais : [en]
BACKGROUND: Glucose control is an important issue in post-operative patients. The objective here was to compare two insulin infusion lines by syringe pumps to assess the impact of medical devices on glycaemic variability ...
Lire la suite >BACKGROUND: Glucose control is an important issue in post-operative patients. The objective here was to compare two insulin infusion lines by syringe pumps to assess the impact of medical devices on glycaemic variability in surgical patients under intensive insulin therapy. This open, prospective, single-centre randomised study was conducted in a fifteen-bed perioperative high dependency unit (HDU) in a university hospital. In total, 172 eligible patients receiving insulin therapy agreed to participate in the study. Subcutaneous continuous glucose monitoring was set up for all patients and an optimised system with a dedicated insulin infusion line for half of the patients. RESULTS: Eighty-six patients were infused via the optimised infusion line and 86 patients via the standard infusion line. No significant difference was found according to the glycaemic lability index score [mean difference between groups (95% CI): -0.09 (-0.34; 0.16), p = 0.49 after multiple imputation]. A glucose control monitoring system indicated a trend towards differences in the duration of hypoglycaemia (blood glucose level below 70 mg dl-1 (3.9 mmol l-1) over 1000 h of insulin infusion (9.7 ± 25.0 h in the standard group versus 4.4 ± 14.8 h in the optimised group, p = 0.059) and in the number of patients experiencing at least one hypoglycaemia incident (25.7 vs. 12.9%, p = 0.052). Time in the target range was similar for both groups. CONCLUSIONS: The use of optimised infusion line with a dedicated insulin infusion line did not reduce glycaemic variability but minimised the incidence of hypoglycaemia events. The choice of the medical devices used to infuse insulin seems important for improving the safety of insulin infusion in perioperative HDU.Lire moins >
Lire la suite >BACKGROUND: Glucose control is an important issue in post-operative patients. The objective here was to compare two insulin infusion lines by syringe pumps to assess the impact of medical devices on glycaemic variability in surgical patients under intensive insulin therapy. This open, prospective, single-centre randomised study was conducted in a fifteen-bed perioperative high dependency unit (HDU) in a university hospital. In total, 172 eligible patients receiving insulin therapy agreed to participate in the study. Subcutaneous continuous glucose monitoring was set up for all patients and an optimised system with a dedicated insulin infusion line for half of the patients. RESULTS: Eighty-six patients were infused via the optimised infusion line and 86 patients via the standard infusion line. No significant difference was found according to the glycaemic lability index score [mean difference between groups (95% CI): -0.09 (-0.34; 0.16), p = 0.49 after multiple imputation]. A glucose control monitoring system indicated a trend towards differences in the duration of hypoglycaemia (blood glucose level below 70 mg dl-1 (3.9 mmol l-1) over 1000 h of insulin infusion (9.7 ± 25.0 h in the standard group versus 4.4 ± 14.8 h in the optimised group, p = 0.059) and in the number of patients experiencing at least one hypoglycaemia incident (25.7 vs. 12.9%, p = 0.052). Time in the target range was similar for both groups. CONCLUSIONS: The use of optimised infusion line with a dedicated insulin infusion line did not reduce glycaemic variability but minimised the incidence of hypoglycaemia events. The choice of the medical devices used to infuse insulin seems important for improving the safety of insulin infusion in perioperative HDU.Lire moins >
Langue :
Anglais
Audience :
Internationale
Vulgarisation :
Non
Établissement(s) :
Université de Lille
CHU Lille
CHU Lille
Collections :
Équipe(s) de recherche :
Innovation/évaluation des médicaments injectables
Innovation/évaluation des dispositifs médicaux de perfusion
Innovation/évaluation des dispositifs médicaux de perfusion
Date de dépôt :
2019-02-26T17:07:08Z
2021-04-12T13:49:14Z
2021-05-28T08:32:31Z
2021-04-12T13:49:14Z
2021-05-28T08:32:31Z
Fichiers
- s13613-017-0298-x.pdf
- Version éditeur
- Accès libre
- Accéder au document