In vitro comparison of two changeover ...
Document type :
Article dans une revue scientifique
DOI :
PMID :
Permalink :
Title :
In vitro comparison of two changeover methods for vasoactive drug infusion pumps: quick-change versus automated relay
Author(s) :
GENAY, Stéphanie [Auteur]
Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Decaudin, Bertrand [Auteur]
Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Lede, Sebastien [Auteur]
Feutry, Frederic [Auteur]
Barthelemy, Christine [Auteur]
Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Lebuffe, Gilles [Auteur]
Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Odou, Pascal [Auteur]
Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]

Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Decaudin, Bertrand [Auteur]

Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Lede, Sebastien [Auteur]
Feutry, Frederic [Auteur]
Barthelemy, Christine [Auteur]
Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Lebuffe, Gilles [Auteur]

Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Odou, Pascal [Auteur]

Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Journal title :
Biomedizinische Technik. Biomedical engineering
Abbreviated title :
Biomed. Eng.-Biomed. Tech.
Volume number :
60
Pages :
377-380
Publication date :
2015
HAL domain(s) :
Sciences du Vivant [q-bio]
English abstract : [en]
This study aimed to compare in vitro two syringe changeover techniques to determine which was better at minimising variations in norepinephrine (NE) delivery: the manual quick-change or automated technique. NE concentration ...
Show more >This study aimed to compare in vitro two syringe changeover techniques to determine which was better at minimising variations in norepinephrine (NE) delivery: the manual quick-change or automated technique. NE concentration was measured continuously using a UV spectrophotometer, and infusion flow rate was monitored by an infusion pump tester. Syringe changeovers were made with either of the two techniques studied. Relays induced disturbances in drug delivery. The temporary increase in NE mass flow rate was significantly higher with manual relays than with automated ones. The automated relay offered a better control of the amounts of NE administered than the quick-change technique.Show less >
Show more >This study aimed to compare in vitro two syringe changeover techniques to determine which was better at minimising variations in norepinephrine (NE) delivery: the manual quick-change or automated technique. NE concentration was measured continuously using a UV spectrophotometer, and infusion flow rate was monitored by an infusion pump tester. Syringe changeovers were made with either of the two techniques studied. Relays induced disturbances in drug delivery. The temporary increase in NE mass flow rate was significantly higher with manual relays than with automated ones. The automated relay offered a better control of the amounts of NE administered than the quick-change technique.Show less >
Language :
Anglais
Audience :
Internationale
Popular science :
Non
Administrative institution(s) :
Université de Lille
CHU Lille
CHU Lille
Collections :
Research team(s) :
Innovation/évaluation des médicaments injectables
Modélisation biopharmaceutique et pharmacocinétique
Innovation/évaluation des dispositifs médicaux de perfusion
Modélisation biopharmaceutique et pharmacocinétique
Innovation/évaluation des dispositifs médicaux de perfusion
Submission date :
2019-02-26T17:07:33Z
2019-07-05T12:02:41Z
2019-07-05T12:02:41Z