Evaluation of cancer drug infusion devices ...
Type de document :
Article dans une revue scientifique: Article original
PMID :
URL permanente :
Titre :
Evaluation of cancer drug infusion devices prior to the implementation of a compounding robot.
Auteur(s) :
Caron, Guillaume [Auteur]
Centre Hospitalier Régional Universitaire [CHU Lille] [CHRU Lille]
Vasseur, Michele [Auteur]
Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Courtin, Justin [Auteur]
Centre Hospitalier Régional Universitaire [CHU Lille] [CHRU Lille]
Masse, Morgane [Auteur]
Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Décaudin, Bertrand [Auteur]
Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
GENAY, Stéphanie [Auteur]
Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Odou, Pascal [Auteur]
Groupe de Recherche sur les formes Injectables et les Technologies Associées (GRITA) - ULR 7365
Simon, Nicolas [Auteur]
Groupe de Recherche sur les formes Injectables et les Technologies Associées (GRITA) - ULR 7365
Centre Hospitalier Régional Universitaire [CHU Lille] [CHRU Lille]
Vasseur, Michele [Auteur]
Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Courtin, Justin [Auteur]
Centre Hospitalier Régional Universitaire [CHU Lille] [CHRU Lille]
Masse, Morgane [Auteur]
Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Décaudin, Bertrand [Auteur]
Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
GENAY, Stéphanie [Auteur]
Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Odou, Pascal [Auteur]
Groupe de Recherche sur les formes Injectables et les Technologies Associées (GRITA) - ULR 7365
Simon, Nicolas [Auteur]
Groupe de Recherche sur les formes Injectables et les Technologies Associées (GRITA) - ULR 7365
Titre de la revue :
Journal of Oncology Pharmacy Practice
Nom court de la revue :
J Oncol Pharm Pract
Pagination :
10781552231170792
Éditeur :
SAGE Publications
Date de publication :
2023-05-20
ISSN :
1477-092X
Discipline(s) HAL :
Sciences du Vivant [q-bio]
Résumé en anglais : [en]
Introduction
Compounding robots are increasingly being implemented in hospital pharmacies. In our hospital, the recent acquisition of a robot (RIVATM, ARxIUM) for intravenous cancer drug compounding obliged us to replace ...
Lire la suite >Introduction Compounding robots are increasingly being implemented in hospital pharmacies. In our hospital, the recent acquisition of a robot (RIVATM, ARxIUM) for intravenous cancer drug compounding obliged us to replace the previously used infusion devices. The objective of the present study was to assess and qualify the new intravenous sets prior to their use in our hospital and prior to the implementation of the compounding robot. Materials and Methods The ChemoLockTM (ICU Medical) was compared with the devices used previously for compounding (BD PhaSealTM, Becton-Dickinson) and infusion (Connect-ZTM, Codan Medical). The connection/disconnection of infusion devices to/from 50 mL infusion bags was tested with a dynamometer (Multitest-i, Mecmesin). Leakage contamination was visualized by a methylene blue assay and was quantified in simulated pump infusions with 20 mg/mL quinine sulfate (N = 36/group); after the analytical assay had been validated, quinine was detected by UV-spectrophotometry at 280 and 330 nm. Groups were compared using chi-squared or Mann–Whitney U tests. Results The connection/disconnection test showed that although all the devices complied with the current standard, there was a statistically significant difference in the mean ± standard deviation compression force (51.5 ± 11.6 for the Connect-ZTM vs. 60.3 ± 11.7 for the ChemoLockTM; p = 0.0005). Leaks were detected in 32 (29.1%) of the 110 tests of the ChemoLockTM. The contamination rates were also significantly different: 13.9% for the BD PhaSealTM versus 75.0% for the ChemoLockTM; p < 0.0001). Discussion/conclusion Our results showed that the new infusion device complied with current standards. However, the presence of contamination emphasizes the need for operators to use the recommended personal protective equipment. Further studies of contamination with cancer drugs are required.Lire moins >
Lire la suite >Introduction Compounding robots are increasingly being implemented in hospital pharmacies. In our hospital, the recent acquisition of a robot (RIVATM, ARxIUM) for intravenous cancer drug compounding obliged us to replace the previously used infusion devices. The objective of the present study was to assess and qualify the new intravenous sets prior to their use in our hospital and prior to the implementation of the compounding robot. Materials and Methods The ChemoLockTM (ICU Medical) was compared with the devices used previously for compounding (BD PhaSealTM, Becton-Dickinson) and infusion (Connect-ZTM, Codan Medical). The connection/disconnection of infusion devices to/from 50 mL infusion bags was tested with a dynamometer (Multitest-i, Mecmesin). Leakage contamination was visualized by a methylene blue assay and was quantified in simulated pump infusions with 20 mg/mL quinine sulfate (N = 36/group); after the analytical assay had been validated, quinine was detected by UV-spectrophotometry at 280 and 330 nm. Groups were compared using chi-squared or Mann–Whitney U tests. Results The connection/disconnection test showed that although all the devices complied with the current standard, there was a statistically significant difference in the mean ± standard deviation compression force (51.5 ± 11.6 for the Connect-ZTM vs. 60.3 ± 11.7 for the ChemoLockTM; p = 0.0005). Leaks were detected in 32 (29.1%) of the 110 tests of the ChemoLockTM. The contamination rates were also significantly different: 13.9% for the BD PhaSealTM versus 75.0% for the ChemoLockTM; p < 0.0001). Discussion/conclusion Our results showed that the new infusion device complied with current standards. However, the presence of contamination emphasizes the need for operators to use the recommended personal protective equipment. Further studies of contamination with cancer drugs are required.Lire moins >