Effectiveness of in-Line Filters to Completely Remove Particulate Contamination During a Pediatric Multidrug Infusion Protocol.
Auteur(s) :
Perez, Maxime [Auteur]
Décaudin, Bertrand [Auteur]
Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Abou Chahla, Wadih [Auteur]
Nelken, Brigitte [Auteur]
Storme, Laurent [Auteur]
Environnement Périnatal et Santé - EA 4489
Environnement périnatal et croissance - EA 4489 [EPS]
Masse, Morgane [Auteur]
Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Barthelemy, Christine [Auteur]
Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Lebuffe, Gilles [Auteur]
Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Odou, Pascal [Auteur]
Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Décaudin, Bertrand [Auteur]

Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Abou Chahla, Wadih [Auteur]
Nelken, Brigitte [Auteur]
Storme, Laurent [Auteur]

Environnement Périnatal et Santé - EA 4489
Environnement périnatal et croissance - EA 4489 [EPS]
Masse, Morgane [Auteur]

Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Barthelemy, Christine [Auteur]

Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Lebuffe, Gilles [Auteur]

Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Odou, Pascal [Auteur]

Groupe de Recherche sur les formes Injectables et les Technologies Associées - ULR 7365 [GRITA]
Titre de la revue :
Scientific Reports
Nom court de la revue :
Sci Rep
Numéro :
8
Pagination :
7714
Date de publication :
2018-05-16
ISSN :
2045-2322
Discipline(s) HAL :
Sciences du Vivant [q-bio]
Résumé en anglais : [en]
The large number of drugs administered simultaneously to neonates and children in hospital results in the formation of particles that are potentially infused. We have investigated the ability of IV in-line filters to ...
Lire la suite >The large number of drugs administered simultaneously to neonates and children in hospital results in the formation of particles that are potentially infused. We have investigated the ability of IV in-line filters to eliminate particulate matter from multidrug infusion lines and so prevent contamination. The impact on particle occurrence of the internal volume of the IV line below the in-line filter was then evaluated. The multidrug therapy given to children was reproduced with and without in-line filtration. Three combinations with a filter were tested to vary the internal volume (V) between the filter and the catheter egress. The catheter was then connected to a dynamic particle count to evaluate the particulate matter potentially administered to children during infusion. The introduction of in-line filters led to a significant reduction in overall particulate matter, from 416,974 [208,479-880,229] to 7,551 [1,985-11,287] particles (p < 0.001). Larger particles of ≥10 and 25 µm were also significantly reduced. Adding an extension set to the egress of the in-line filter (V = 1.7 mL) caused a significant increase in particulate contamination for both. This study showed that in-line filtration is an effective tool in preventing particle administration to patients. Their position in the infusion in-line is therefore important because of its impact on internal volume and drug particle formation.Lire moins >
Lire la suite >The large number of drugs administered simultaneously to neonates and children in hospital results in the formation of particles that are potentially infused. We have investigated the ability of IV in-line filters to eliminate particulate matter from multidrug infusion lines and so prevent contamination. The impact on particle occurrence of the internal volume of the IV line below the in-line filter was then evaluated. The multidrug therapy given to children was reproduced with and without in-line filtration. Three combinations with a filter were tested to vary the internal volume (V) between the filter and the catheter egress. The catheter was then connected to a dynamic particle count to evaluate the particulate matter potentially administered to children during infusion. The introduction of in-line filters led to a significant reduction in overall particulate matter, from 416,974 [208,479-880,229] to 7,551 [1,985-11,287] particles (p < 0.001). Larger particles of ≥10 and 25 µm were also significantly reduced. Adding an extension set to the egress of the in-line filter (V = 1.7 mL) caused a significant increase in particulate contamination for both. This study showed that in-line filtration is an effective tool in preventing particle administration to patients. Their position in the infusion in-line is therefore important because of its impact on internal volume and drug particle formation.Lire moins >
Langue :
Anglais
Audience :
Internationale
Vulgarisation :
Non
Établissement(s) :
Université de Lille
CHU Lille
CHU Lille
Collections :
Équipe(s) de recherche :
Innovation/évaluation des médicaments injectables
Modélisation biopharmaceutique et pharmacocinétique
Innovation/évaluation des dispositifs médicaux de perfusion
Modélisation biopharmaceutique et pharmacocinétique
Innovation/évaluation des dispositifs médicaux de perfusion
Date de dépôt :
2019-02-26T17:06:55Z
2019-07-05T11:53:23Z
2021-05-19T06:51:50Z
2019-07-05T11:53:23Z
2021-05-19T06:51:50Z